Mostrar el registro sencillo del ítem
‘‘Determinación de la tasa de referencia de kerma en aire para una fuente de Ir-192 y Validación del sistema de planificación Brachyvision para braquiterapia de alta tasa de dosis’’
| dc.contributor.advisor | Saba Effio, Augusto | es_PE |
| dc.contributor.advisor | Mori Rojas, Juan José | es_PE |
| dc.contributor.author | Rivera Barco, Jerry Astrid | es_PE |
| dc.date.accessioned | 2026-09-14T15:22:38Z | |
| dc.date.available | 2026-09-14T15:22:38Z | |
| dc.date.issued | 2026-09-02 | |
| dc.identifier.uri | https://hdl.handle.net/20.500.12893/17707 | |
| dc.description.abstract | La braquiterapia de alta tasa de dosis (HDR) es una técnica esencial en el tratamiento oncológico, cuyo desempeño depende de factores clínicos y físicos, entre ellos la precisión dosimétrica en la entrega de la dosis y la adecuada validación del sistema de planificación. En este estudio se plantea la determinación de la tasa de referencia de kerma en aire para una fuente de Ir- 192 y la validación del sistema BrachyVision utilizado en la Clínica ONCORAD, sede Trujillo, con el propósito de asegurar la calidad y seguridad en los tratamientos. La investigación se desarrolla bajo un enfoque cuantitativo, de tipo experimental y orientada a la aplicación en dosimetría clínica. Se realiza el control de calidad del conjunto dosimétrico empleado: cámara de pozo, electrómetro, barómetro y termómetro, siguiendo las recomendaciones internacionales establecidas en el IAEA TECDOC-1274 y los fundamentos del manual Braquiterapia de Alta Tasa de Dosis para Físicos (INCA). Posteriormente, se determina la tasa de referencia de kerma en aire de la fuente mediante la metodología descrita en el TECDOC- 1274 y se efectúa la verificación del sistema BrachyVision, comparando los valores calculados por el sistema de planificación con los datos de referencia proporcionados por el fabricante, bajo condiciones controladas. Este trabajo busca contribuir al fortalecimiento del aseguramiento de la calidad en braquiterapia HDR, proporcionando un procedimiento estandarizado para la determinación y validación dosimétrica, que favorezca la trazabilidad institucional y la confiabilidad en la planificación de tratamientos. | es_PE |
| dc.description.abstract | High-dose-rate (HDR) brachytherapy is a key modality in cancer treatment, whose effectiveness relies on both clinical and physical factors, particularly the dosimetric accuracy in dose delivery and the proper validation of the treatment planning system. This study focuses on determining the reference air-kerma rate of an Ir-192 source and validating the BrachyVision system used at Clínica ONCORAD, Trujillo, with the aim of ensuring the quality and safety of treatments. The research follows a quantitative, experimental approach and is applied to clinical dosimetry. Quality control of the dosimetric system—well chamber, electrometer, barometer, and thermometer—is performed according to international recommendations from the IAEA TECDOC-1274 and the guidelines described in the High-Dose-Rate Brachytherapy Physics Manual (INCA). Subsequently, the reference air-kerma rate of the source is determined following the TECDOC-1274 methodology, and the BrachyVision system is verified by comparing the calculated planning values with the reference data provided by the manufacturer under controlled conditions. This work contributes to quality assurance in HDR brachytherapy by establishing a standardized procedure for dosimetric determination and validation, strengthening institutional traceability and improving confidence in treatment planning. | es_PE |
| dc.format | application/pdf | es_PE |
| dc.language.iso | spa | es_PE |
| dc.publisher | Universidad Nacional Pedro Ruiz Gallo | es_PE |
| dc.rights | info:eu-repo/semantics/openAccess | es_PE |
| dc.rights.uri | http://creativecommons.org/licenses/by-nc-sa/4.0/ | es_PE |
| dc.subject | Braquiterapia | es_PE |
| dc.subject | Radioterapia | es_PE |
| dc.subject | Iridio-192 | es_PE |
| dc.title | ‘‘Determinación de la tasa de referencia de kerma en aire para una fuente de Ir-192 y Validación del sistema de planificación Brachyvision para braquiterapia de alta tasa de dosis’’ | es_PE |
| dc.type | info:eu-repo/semantics/bachelorThesis | es_PE |
| thesis.degree.name | Licenciado en Física | es_PE |
| thesis.degree.grantor | Universidad Nacional Pedro Ruiz Gallo - Facultad de Ciencias Físicas y Matemáticas | es_PE |
| thesis.degree.discipline | Física | es_PE |
| dc.type.version | info:eu-repo/semantics/submittedVersion | es_PE |
| dc.publisher.country | PE | es_PE |
| dc.subject.ocde | http://purl.org/pe-repo/ocde/ford#1.03.04 | es_PE |
| renati.author.dni | 77478881 | |
| renati.advisor.dni | 16600649 | |
| renati.advisor.orcid | https://orcid.org/0000-0002-3602-2274 | es_PE |
| renati.type | http://purl.org/pe-repo/renati/type#tesis | es_PE |
| renati.level | http://purl.org/pe-repo/renati/nivel#tituloProfesional | es_PE |
| renati.discipline | 533056 | es_PE |
| renati.juror | Cieza Ramos, Arnulfo | es_PE |
| renati.juror | Rivera Vásquez, Juan Pedro | es_PE |
| renati.juror | Prada Marchena, Jara Hernán | es_PE |
Ficheros en el ítem
Este ítem aparece en la(s) siguiente(s) colección(ones)
-
Física [67]